|
ACPD
in Detail
Practical Deployment
The ACPD technique requires two connections between the instrument
and the inspection site - namely the current output, and the voltage
input (from a single probe for inspection and sizing, or from one
of a number of probes during crack growth monitoring). This section
describes the general rules for deployment on simple specimens such
as flat plates or uniform cylinders. Deployment on more complex
geometries is described in the next section.
Current Input
The current is generally injected into the specimen by direct contact.
The two contacts should be placed equidistant on either side of
the crack site and with the crack lying perpendicular to the line
between the contacts (see Figure 1). The contacts should be about
150mm away from the crack, if possible. The further apart the contacts
are the more uniform the field is. On ferritic steel, the field
contacts are most conveniently attached by magnets. On non-ferrous
materials they have to be fixed on mechanically (e.g. screwed onto
spot-welded studs, spot welded directly to the surface, or attached
to suitable points with crocodile clips).

Figure 1 - Injected current input on a flat plate.
In certain situations, such as for axial cracks on a cylinder,
an injected field is not very practical. In this case it is better
to use the field leads to induce a current into the specimen by
deploying them above the specimen surface perpendicular to the crack
(see Figure 2).
For other situations where it it not possible to inject a current,
but where it is also difficult to lay inducing wires, special purpose
ACPD voltage probes can be produced which carry a self-contained
induction coil.

Figure 2 - Induced field input for axial crack in
cylinder (cut-away view).
Voltage Probe Connection
It is important to maintain good electrical contact with the sample
with both pins for several seconds while the voltage display settles.
For this reason, probe pins are usually sprung so that contact is
maintained even if the probe is moved slightly. Such probes should
be held down with the springs fully compressed to minimise the area
of loop between the pins. If stable readings cannot be obtained,
it is normally because the electrical contact is not good enough.
The metal surface must be cleaned to bight metal.
Taking Measurements
In its simplest deployment, the ACPD technique requires three parameters
in order to produce a crack depth; these area voltage reading with
probe straddling the crack, a voltage reading with the probe adjacent
but not straddling the crack, and the spacing between the voltage
probe contacts. The voltage readings are in arbitrary units because
only the ration of the two is used.
If the position of a crack is not known, the voltage probe should
be moved along a line perpendicular to the expected crack edge.
The display should be watched fro any sudden changes in reading.
The reading will jump up from one steady value to another higher
value when the leading probe contact crosses the crack. Not that
these values should be stable, repeatable and non-zero; non-repeatable
or very low readings indicate loss of electrical contact by either
the voltage probe or the field input.
Having determined the crack location, the two required voltages
should be obtained as shown in Figure 3. The probe should be edged
slowly towards the crack until the voltage reading jumps up. This
value should be recorded as the cross-crack voltage, Vc. Readings
should be observed on the instrument front panel for stability.
A reading should be stable within +/- 1 digit for several seconds
before being captured. The captured value should be checked to ensure
it matches the value observed on the instrument before capture.
The probe should then be slowly backed off until the voltage reading
drops down again. This value should be recorded as the reference
voltage, Vr. The purpose of this procedure is to obtain voltage
readings as close together as possible to minimise the effect of
any variation i input field strength. Such variations occur, even
with a nominally uniform input field, around short cracks or cracks
in materials with a large skin depth.

Figure 3 - Accurate measurement of ACPD voltages.
When use on flat plate, the spacing between the voltage probe contacts
along the metal surface will be the same as the straight-line separation
in air. This can then be measured with a ruler. On any other surface,
the separation along the metal surface will be longer and should
be measured directly, using flexible tape, for example, at the site
where the readings are taken.
The voltage changes shown in Figure 3 apply for a long uniform
depth crack in a thin-skin material. If the crack is short compared
to its depth (i.e. length < 5 x depth), or is in a thick-skin
material, the voltage is not constant either side of a crack - instead
the distribution contains sharp cusps (as shown in Figure 4). In
such cases, it is important to measure the voltages at the cusp
points (as close t the crack edges as possible) otherwise the crack
depth will be underestimated. In cases where it is not possible
to measure the cusp voltages (such as when monitoring crack growth
with fixed probes) it is possible to correct for this underestimate
if the separation between the crack and the nearest voltage contact
is known (refer to TSC for details).

Figure 4 - Comparison
of ACPD voltages from a probe crossing (left) a long and (right)
a short crack.
Closing Comments
As long as care is taken over field input placement
and probe deployment as described above, ACPD is a very easy technique
to use to obtain accurate crack sizing in a wide range of applications.
Over the last twenty years, staff at TSC and University College
London have built up a wealth of experience in the successful use
of ACPD, and will be pleased to help with any applications not covered
in this document.
ACPD
Measurement / Introduction /
ACPD In Detail / Practical Deployment 1 2
3 Next
|